PHOIBOS 150 NAP 1D-DLD
Mounting
Mounting Flange

DN150 CF (8" OD)

Working Distance

300 - 500 µm (for standard nozzle with 300 µm diameter)

Operation
Energy Dispersion

Hemisphere

Slits/Apertures

8 Entrance, 3 Exit slits and Iris aperture

Magnetic Shielding

Double µ-Metal Shielding

Lens Modes

Angular Resolving and Transmission Lens Modes

Kinetic Energy Range

0 - 3500 eV

Pass Energies

0 - 550 eV continously adjustable

Detector

1D-DLD Detector

Measurement Modes

Snapshot mode, Sweeping mode, Fixed energy mode

Energy Window

13% of Pass Energy

Electric Isolation

up to 7 keV

Electronics

HSA 3500 plus HT 100 for analyzer and HSA 3500 plus HT 173 for pre-lens

Working Pressure

up to 30 mbar (higher pressure values achievable with corresponding nozzle diameters and differential pumping packages)

Performance
Acceptance Angle

±22°

Angular Resolution

only with 2D detector

Energy Resolution

< 10 meV

Lateral Resolution

only with 2D detector

Smallest Acceptance Spot

Without imaging lens module the smallest acceptance spot is defined by the excitation spot of X-ray source or nozzle diameter

Detector Channels

Max. 960 energy channels (240 and 120 channels for binning 4 and 8, respectively)

XPS Count Rates UHV

70 kcps (guaranteed), 150 kcps (achievable) *

XPS Count Rates 10 mbar

7 kcps (guaranteed), 15 kcps (achievable) *

XPS Count Rates 25 mbar

0.5 kcps (guaranteed), 1.5 kcps (achievable) *

.

* Ag 3d, FWHM < 0.85 eV, small spot monochromated X-ray source µ-FOCUS 600, Al Kα anode, 20 W, Spot size < 250 µm